INCUCELL 22
Incubators with Natural Forced Circulation
-Type ECO Line MC000220 -Type EVO Line MC001906 Suitable for safe processing of microbiologic cultures. The INCUCELL® line is characterised by noiseless operation and slow air flow in usable space. The INCUCELL® V line (version with fan) guarantees faster and more precise temperature distribution with small variations. Special fields of use include mainly biologic and microbiologic laboratories. They can be also used for quality tests in pharmaceutical industry and cosmetics, for tests in veterinary medicine and food processing industry.- natural, accurate – laboratory incubator
- chamber volume 22 litres
- temperature range from 5°C above ambient to 100°C
- characterized by noiseless operation and fine air flow in the usable space
- high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
- high-quality and safe drying, heating and tempering of common laboratory materials in laboratory and industrial areas, suitable for safe processing of microbiological cultures
- standard version with microprocessor control system Fuzzy logic, controls placed on a foil keyboard, process information displayed on LED display
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Specification
Interior of stainless steel material No. 1.4301
volume | 22 l |
depth | 350 mm |
width | 240 mm |
height | 300 mm |
Tray
standard equipment | 2 pcs |
storage area (w × d) | 185 × 265 mm |
racks | 4 max. number |
distance between guide rails | 60 mm |
Maximal weight of the load*)
inside the oven | 25 kg/oven |
per tray | 10 kg/tray |
Door
1 pcs |
External dimensions (including door and handle)
depth | 560 mm |
diameter of the air branch outer / inner | 52/49 mm |
width | 406 mm |
610 (N) |
Package dimensions (three layers carton)
depth | 720 mm |
width | 500 mm |
depth | 810 mm |
Weight
brut | 36 kg |
nett | 31 kg |
Working temperature (beginning of the regulation)
from 5 °C above ambient temp. to: | 100 °C |
Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door
time variation cca (±) | 0,4 °C |
space deviation cca (±) | 1,1 °C |
Time required to reach 37 °C with closed air flap and voltage 230 V
70 min |
Heat emission at 37 °C
20 W |
Air exchange speed at 37 °C
4 per hour |
Electricity – mains 50/60 Hz
max. power | 0,24 kW |
nominal voltage | 230 V |
current | 1,1 A |
power by stand by | 5 W |