VENTICELL 22
Ovens with Forced Circulation
Type Eco Line MC000210
Type Evo Line MC001902
Due to a patented ventilation system the air within the VENTICELL® chamber is ventilated in a regular spiral way. This leads to a homogenous temperature profile throughout the chamber and short heating times. Operating economy is ensured by higher rate and precision of heating in laboratories. Especially suitable for very moist goods.- precise and fast laboratory drying oven
- chamber volume 22 litres
- temperature range from 10°C above ambient to 250/300°C
- forced air circulation system in the chamber guarantees homogeneous temperature distribution in all the processes of drying, heating and sterilization of materials in laboratories
- high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
- higher speed and accuracy of all the tempering procedures ensures economical operation, especially suitable for high humidity materials, for demanding and accurate tests and processes
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Specification
Interior of stainless steel material No. 1.4301
volume | 22 l |
depth | 320 mm |
width | 240 mm |
height | 300 mm |
Tray
standard equipment | 2 pcs |
storage area (w × d) | 185 × 265 mm |
racks | 4 max. number |
distance between guide rails | 60 mm |
Maximal weight of the load *)
inside the oven | 25 kg/oven |
per tray | 10 kg/tray |
Number of doors
1 pcs |
External dimensions (including door and handle)
depth | 560 mm |
diameter of the air branch outer/inner | 52/49 mm |
width | 406 mm |
height (inclusive foots – N) | 610 (N) mm |
Package dimensions (three layers carton)
depth | 720 mm |
width | 500 mm |
depth | 810 mm |
Weight
brut | 36 kg |
net | 31 kg |
Working temperature **) (beginning of the regulation) – VENTICELL / VENTICELL plus
from 10 °C above ambient temp. to | 250 °C |
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door, > 50 °C
time variation cca (±) | 0,3 °C |
space deviation (±) % cca (±)% of the reached temperature | 1,1 % |
Time required to reach 250 °C with closed air flap and voltage 230 V
28 min |
Heat emission at 250 °C
420 W |
Air exchange speed at 250 °C
45 cca/hour |
Electricity – mains 50/60 Hz
max. power input | 0,96 kW |
nominal voltage | 230 V |
current | 4,2 A |
stand by | 5 W |